Repositório Digital

A- A A+

Mercury speciation in urban landfill leachate by cold vapor generation atomic absorption spectrometry using ion exchange and amalgamation

.

Mercury speciation in urban landfill leachate by cold vapor generation atomic absorption spectrometry using ion exchange and amalgamation

Mostrar registro completo

Estatísticas

Título Mercury speciation in urban landfill leachate by cold vapor generation atomic absorption spectrometry using ion exchange and amalgamation
Autor Dressler, Valderi L.
Pozebon, Dirce
Flores, Éder L.M.
Paniz, José Neri Gottfried
Flores, Erico Marlon de Moraes
Resumo Foi desenvolvido um sistema em linha para a especiação de Hg orgânico (Hgorg) e Hg inorgânico (Hginorg) em chorume. Clorocomplexos produzidos a partir do Hginorg são separados em uma coluna trocadora de ânions. O Hg retido na coluna é eluído com NaBH4 na presença de HCl. O Hg0 produzido é medido por espectrometria de absorção atômica com geração de vapor frio (CV AAS). O Hgorg, que não fica retido na coluna, é misturado com NaBH4 produzindo espécies voláteis de Hg, as quais são amalgamadas numa rede de Au/Pt. O Hg é removido desta coluna por aquecimento a 600 °C e medido por CV AAS. Para a especiação de Hginorg e Hgorg em chorume, as amostras são centrifugadas, filtradas e tratadas com HCl. Os limites de detecção (3 s) são 9 e 12 ng L-1 de Hg para Hginorg e Hgorg, respectivamente. A freqüência de análise é 22 h-1
Abstract This work deals with on line speciation of organic Hg (Hgorg) and inorganic Hg (Hginorg) in urban landfill leachate. Chlorocomplexes of Hginorg are produced in HCl medium and retained on an anion exchange column. The retained Hg is eluted with NaBH4 in presence of HCl. Hg0 is produced and then measured by cold vapor atomic absorption spectrometry (CV AAS). Hgorg does not interact with the resin and is mixed with NaBH4 at the exit of the column producing volatile species of Hg that are then amalgamated by a gauze of Au/Pt. Mercury is removed from the gauze by heating to 600 °C and subsequently measured by CV AAS. For Hginorg and Hgorg speciation in urban landfill leachate, the samples are centrifuged, filtered and treated with HCl. The detection limits (3 s) of Hginorg and Hgorg are 9 and 12 ng L-1 of Hg, respectively. The analysis frequency is 22 h-1.
Contido em Journal of the Brazilian Chemical Society. São Paulo. Vol. 20, n. 9(2009), p. 1659-1666
Assunto Aterro sanitário
Chorume : Compostos orgânicos
Espectrometria de absorção atômica
Mercurio
[en] Amalgamation
[en] CVG AAS
[en] Flow analysis
[en] Ion exchange
[en] Landfill leachate
[en] Mercury speciation
Origem Nacional
Tipo Artigo de periódico
URI http://hdl.handle.net/10183/24599
Arquivos Descrição Formato
000733622.pdf (424.1Kb) Texto completo (inglês) Adobe PDF Visualizar/abrir