dc.description.abstractThis work purpose is to create a positioning automated system of a tailstock to perform impact resistance tests on silicone mammary implants. This system is capable to measure and oversight the positioning through interrelation between three main components: programmable logic controller, human-machine interface and magnetostrictive transducer. Together, these components form an operational closed loop that ensures an appropriate positioning for the impact device. The paper describes how the closed loop works and also the algorithm implemented in the programmable logic controller which surveys the positioning. As a last topic, the paper presents the operator work on the machine’s operation in conjunction with human-machine interface. The results were satisfactory and in accordance with the limits determined on ABNT NBR ISO 14607 for this method of tests.en
dc.format.mimetypeapplication/pdfpt_BR
dc.language.isoengpt_BR
dc.relation.ispartofJournal of physics: conference series [recurso eletrônico]. Bristol, United Kingdom. Vol. 648, conference 1 (2015), 7 p.pt_BR
dc.rightsOpen Accessen
dc.subjectControlador programávelpt_BR
dc.subjectTransdutores ultrasônicospt_BR
dc.subjectEnsaios de impactopt_BR
dc.titleVertical positioning surveillance by magnetostrictive transducerpt_BR