dc.description.abstractA determinação da vazão aduzida em sistemas de condutos forçados, por exemplo, em uma usina hidrelétrica, é uma tarefa difícil de ser executada. Isto se deve a fatores como as dimensões físicas da instalação, as dificuldades de implementação de sensores no sistema e a existência de trechos fisicamente adequados para se executar a medição, etc. Dentre os diversos métodos existentes para a medição de grandes vazões o método de Pitot apresenta uma série de vantagens. Entretanto, a medição depende da calibragem do equipamento, que nem sempre é uma tarefa fácil de ser executada. Neste trabalho, apresenta-se uma metodologia para a calibragem de tubos de Pitot utilizando-se anemometria LASER. Assim, descreve-se o princípio de funcionamento do tubo de Pitot, a bancada de testes utilizada para a realização dos ensaios, incluindo os principais acessórios que permitem a variação da vazão na seção de teste, o anemômetro LASER Doppler utilizado e o arranjo físico adotado nos ensaios de calibragem do tubo de Pitot, que no caso foi o Pitot Cole. Com os dados obtidos foi possível determinar o coeficiente do Tubo de Pitot. Foi possível também fazer o levantamento da influência do bocal do equipamento no campo de velocidade em volta do tubo de Pitot. Posteriormente, fez-se uma análise comparativa da expressão obtida nos ensaios com outra expressão extraída da literatura. Apresenta-se a metodologia desenvolvida para a análise das incertezas das medições de velocidades e vazões determinadas em laboratório.pt_BR
dc.description.abstractTo measure the inlet flow in closed systems, such as the inlet flow of a turbine in an electrical power plant, is a typically difficult task. There are several reasons for this, ranging from the size of the piping involved to the difficulty of finding an adequate position to install measuring devices (sensors) and the installation itself. Among several flow measurement methods, the Pitot tubes present a number of advantages. Calibration of this device, however, is far from trivial. The main goal of this work is to describe a framework to calibrate Pitot tubes using a LASER Doppler anenometer as a standard. The aforementioned description details the main parts of the laboratory setup and also the procedures followed during the calibration tests. The results discussed concern a Cole Pitot tube. The main coefficient of this device was estimated from test data. A major concern throughout the work was how the presence of the Pitot tube would influence the velocity vector field in the neighborhood of the stagnation pressure point. The results obtained are compared and discussed in the light of well-known results established in the literature. Preliminary results on confidence intervals for the measured velocity and flow are also presented and discussed.en
dc.format.mimetypeapplication/pdfpt_BR
dc.language.isoporpt_BR
dc.relation.ispartofRbrh: revista brasileira de recursos hídricos. Porto Alegre,RS: ABRH. Vol. 9, n. 1(jan./mar. 2004), p. 113-126pt_BR
dc.rightsOpen Accessen
dc.subjectFlow measurementen
dc.subjectVazões : Mediçãopt_BR
dc.subjectPitot tubesen
dc.subjectHydroelectric power plantsen
dc.titleMetodologia para calibração de tubo de Pitot Cole utilizando anemometria LASERpt_BR
dc.title.alternativeCalibration method for a cole pilor tube using Laser anemometry en